Gt2-500, Series instruction manual, High-accuracy digital contact sensor
KEYENCE GT2-H12xxF User Manual. High-precision Digital Contact Sensor. Instruction Manual. Read this instruction manual carefully prior to operating this product. For connection, settings and precautions on use, make sure to read the GT2 Series instruction manual included with the amplifier. XG-X Series Users Manual (Alignment Function Edition) Available for Purchasers Only. Configuration and adjustment via PC. XG-X Series XG-X VisionEditor Reference Manual Available for Purchasers Only. Capturing/saving images and results to a PC. XG-X VisionTerminal Remote Control Software for XG-X Series User's Manual. KEYENCE on Innovations with Ed Begley, Jr. Full SegmentKEYENCE GT-71A/GT-H10 In general contact type digital sensor amplifier / head 動作確認 Keyence Gt2 Series User Manual The GT2 Series is available in a wide range of heads, including pencil, air, and low-speed pressure types. Choose a design most suitable for the application.
1
GT2-500-M-E
0
High-accuracy Digital Contact Sensor
GT2-500
Latest deep exploration 6.5 keygen 2016 and full version 2016. Series Instruction Manual
■
Symbols
The following symbols alert you to matters concerning the prevention of
human injury and product damage. Be sure to read these messages carefully.
Keyence Lvdt Gt2
Safety Information for GT2-500 Series
■
Other precautions
● Effects of surrounding air temperature
To use the GT2-500 Series with high accuracy, do not use the GT2-500 Series
in an environment in which the surrounding air temperature changes sharply.
It will take about 40 minutes for the 12 mm type, and 60 minutes for the 32 mm/
50 mm type for the internal temperature to stabilize when the surrounding air
temperature has changed by 10°C.
● Warming up
The circuit is not stable immediately after the power turns ON, which
sometimes causes the indicated value to gradually fluctuate. Wait for about
5 minutes when using the 12 mm type and about 10 minutes when using
the 32/50 mm type after the power turns on before starting operation.
● GT2-PA12K/PA12 (See illustration 1 for the pneumatic circuit)
•
This model uses a single-acting cylinder. The internal spring will
return the sensor head to its home position.
•
The air supply hole cannot be removed.
•
Use an air filter, mist separator, etc. to provide clean dry air. Empty the
drainage from the filter regularly, before it exceeds the specified line.
•
Before connecting the air tube to the air supply hole, be sure to blow plenty of
air through the pipes (flushing) to remove any foreign matter.
•
Make sure that the air pressure of the supplied air is constant and in the range
of 0.24 to 0.26 MPa. Use a precision regulator to control the air pressure. If
the air pressure is below 0.24 MPa, the spindle may not extend fully.
•
When supplying air (with the spindle extended), up to 3L/min of air will be
emitted from the tip of the dust seal.
•
The measuring ability of the device changes according to the air pressure
of the air supply. Refer to
'Specifications' (page 9) for details.
● GT2-A12K(L)/A12(L)/A32/A50 (See illustration 1 for the pneumatic circuit)
•
This model uses a single-acting cylinder. The internal spring will
return the sensor head to its home position when air pressure is
removed. DO NOT supply air to the exhaust valve.
•
The air cylinder cannot be removed.
•
Use the regulator to supply the stable air pressure to the sensor head.
•
Use the air filter and mist separators to supply dry air. Empty the drainage
from the filter regularly, before it exceeds the specified line.
•
Make sure that the air pressure of the air supply is constant and in
the range of 0.25 MPa to 0.5 MPa.
•
The coupling socket and the exhaust valve that are supplied with the
air cylinder are dedicated for this product. They cannot be removed.
● Power supply
•
Noise superimposed on the power supply may cause malfunction. Be sure to
use the DC stabilized power supply provided with an insulation transformer.
•
In the case of a commercially available switching regulator, be
sure to ground the frame ground terminal or the ground terminal.
● About dust boot
When the dust boot of the sensor head is damaged, use an optional dust
boot sold separately (For GT2-P12K(F)/P12(F)/H12K(F)/H12(F)/A12K/
A12: OP-84332 (Material: NBR (attached when shipped)). For GT2-
P12K(F)/P12(F): OP-87859 (Material: Fluororubber). For GT2-H32/A32:
OP-84459 (Material: NBR (attached when shipped)). For GT2-H50/A50
(Material: NBR (attached when shipped)).) Attach the dust boot correctly.
The dust boot may deteriorate, depending on the environment it is
used in. In this case, replace the dust boot regularly.
* Do not use the dust boot on the GT2-P12KL/P12L/PA12K/PA12/
H12KL(F)/H12L(F)/H32L.
Precautions on Regulations and Standards
■
CE Marking
● EMC Directive (2004/108/EC)
•
Applicable standard EMI: EN61326-1, Class A
EMS: EN61326-1
•
This product is designed for use in industrial environments.
Remarks This specification does not give any guarantee that the end-product
with this product incorporated complies with the essential requirements of EMC
Directive. The manufacturer of the end-product is solely responsible for the
compliance on the end-product itself according to EMC Directive.
WARNING
It indicates a hazardous situation which, if not avoided,
could result in death or serious injury.
NOTICE
It indicates a situation which, if not avoided, could result in
product damage as well as property damage.
WARNING
If the following conditions are encountered, immediately
turn OFF the power. Continuing to use the GT2-500
Series under these abnormal conditions may cause
equipment failure.
• When water or foreign matter enters the controller
• When the GT2-500 Series is dropped or the housing is
damaged
• When the GT2-500 Series produces smoke or an
abnormal smell
WARNING
• Do not use the GT2-500 Series with a voltage other
than specified voltage, as this may cause fire, electric
shock or equipment failure.
• Do not disassemble or modify the GT2-500 Series.
This may cause fire or electric shock.
• Do not use this product for the purpose to protect a
human body or a part of human body.
• This product is not intended for use as explosion-proof
product. Do not use this product in a hazardous location
and/or potentially explosive atmosphere.
NOTICE
• Be sure to turn OFF the power of the GT2-500 Series and
any connected devices before connecting or disconnecting
the cables. Otherwise, there may be a risk of damage.
• Do not turn OFF the power while setting parameters.
Otherwise, the settings may be partially or completely lost.
• At startup and during operation, be sure to monitor the
function performance of the GT2-500 Series.
• Do not modify the GT2-500 Series or use it in any way
other than described in the specifications.
• The function performance of products used or modified
in this way cannot be guaranteed.
• When the GT2-500 Series is used in combination with
other instruments, function performance may be degraded,
depending on operating conditions and the surrounding
environment. Use the GT2-500 Series after fully studying
the effect of combined use with other instruments.
• Do not expose the GT2-500 Series and peripheral
devices to sudden temperature change. This may
cause condensation, damaging the equipment.
● Environmental conditions
To use the GT2-500 Series properly and safely, do not
install the GT2-500 Series in the following locations.
Use of this equipment in an improper environment
may cause equipment failure.
•
● About dust boot
When the dust boot of the sensor head is damaged, use an optional dust
boot sold separately (For GT2-P12K(F)/P12(F)/H12K(F)/H12(F)/A12K/
A12: OP-84332 (Material: NBR (attached when shipped)). For GT2-
P12K(F)/P12(F): OP-87859 (Material: Fluororubber). For GT2-H32/A32:
OP-84459 (Material: NBR (attached when shipped)). For GT2-H50/A50
(Material: NBR (attached when shipped)).) Attach the dust boot correctly.
The dust boot may deteriorate, depending on the environment it is
used in. In this case, replace the dust boot regularly.
* Do not use the dust boot on the GT2-P12KL/P12L/PA12K/PA12/
H12KL(F)/H12L(F)/H32L.
Precautions on Regulations and Standards
■
CE Marking
● EMC Directive (2004/108/EC)
•
Applicable standard EMI: EN61326-1, Class A
EMS: EN61326-1
•
This product is designed for use in industrial environments.
Remarks This specification does not give any guarantee that the end-product
with this product incorporated complies with the essential requirements of EMC
Directive. The manufacturer of the end-product is solely responsible for the
compliance on the end-product itself according to EMC Directive.
WARNING
It indicates a hazardous situation which, if not avoided,
could result in death or serious injury.
NOTICE
It indicates a situation which, if not avoided, could result in
product damage as well as property damage.
WARNING
If the following conditions are encountered, immediately
turn OFF the power. Continuing to use the GT2-500
Series under these abnormal conditions may cause
equipment failure.
• When water or foreign matter enters the controller
• When the GT2-500 Series is dropped or the housing is
damaged
• When the GT2-500 Series produces smoke or an
abnormal smell
WARNING
• Do not use the GT2-500 Series with a voltage other
than specified voltage, as this may cause fire, electric
shock or equipment failure.
• Do not disassemble or modify the GT2-500 Series.
This may cause fire or electric shock.
• Do not use this product for the purpose to protect a
human body or a part of human body.
• This product is not intended for use as explosion-proof
product. Do not use this product in a hazardous location
and/or potentially explosive atmosphere.
NOTICE
• Be sure to turn OFF the power of the GT2-500 Series and
any connected devices before connecting or disconnecting
the cables. Otherwise, there may be a risk of damage.
• Do not turn OFF the power while setting parameters.
Otherwise, the settings may be partially or completely lost.
• At startup and during operation, be sure to monitor the
function performance of the GT2-500 Series.
• Do not modify the GT2-500 Series or use it in any way
other than described in the specifications.
• The function performance of products used or modified
in this way cannot be guaranteed.
• When the GT2-500 Series is used in combination with
other instruments, function performance may be degraded,
depending on operating conditions and the surrounding
environment. Use the GT2-500 Series after fully studying
the effect of combined use with other instruments.
• Do not expose the GT2-500 Series and peripheral
devices to sudden temperature change. This may
cause condensation, damaging the equipment.
● Environmental conditions
To use the GT2-500 Series properly and safely, do not
install the GT2-500 Series in the following locations.
Use of this equipment in an improper environment
may cause equipment failure.
•
Locations outside
•
Locations with high humidity, a large amount of dust,
or poor ventilation
•
Locations where the temperature rises excessively
due to direct sunlight, etc.
•
Locations near corrosive or flammable gas
•
Locations where the GT2-500 Series is directly
subjected to vibration or impact
•
Locations where water, oil or chemicals may come
into contact with the GT2-500 Series
•
Locations where static electricity may easily occur
● Noise countermeasures
Installation near a source of electrical noise such as a power
source or a power cable may cause malfunction or failure of
the equipment. Adopt appropriate countermeasures against
noise by using a noise filter or wiring cables in separate ducts,
attaching insulation to the amplifier or the sensor head, etc.
NOTICE
● Handling of the sensor head
•
The GT2-500 Series and peripheral devices are precision
machines. Do not drop, or cause any other impact to these
devices. Doing so may cause damage or malfunction.
•
Do not apply weight greater than what is listed below to the spindle.
Do not apply rotational torque. Otherwise the spindle may break.
•
The GT2-P12KL(F)/P12L(F)/PA12K/PA12 spindle can rotate a
maximum of about 4.5°. When using an offset contact (OP-
77683) or similar, if pressure in the rotational direction is
applied to the contact, the measurement position may change.
Although the GT2-PA12K/PA12/H12K/H12KF/H12/H12F/
H32/H50/A12K/A12/A32/A50 has a protection rating of
IP67, avoid using it immersed in water or in places where
liquid such as oil may come into contact with it.
•
Although the GT2-P12K(F)/P12(F) has a protection rating of
IP67G/NEMA Type 13, some types of oil may damage the device.
•
If dust, metal powder, or similar becomes attached to the
sliding part of the sensor head, mechanical responses may
become slow. If this happens, replace the dust seal with a
replacement dust seal (OP-87932). Refer to the instruction
manual included with the replacement dust seal (OP-
87932) for instructions on how to replace the dust seal.
30 N
100 N
1 N
AIR
Mist separator
Pressure source
Air filter
Solenoid
valve
Precision
regulator
Air supply part
Speed
controller
Illustration 1 - Pneumatic circuit
96M12757